Pumps
High-pressure pumps are needed to force solvents through packed stationary phase beds. Smaller bed particles (e.g. 3 μm) require higher pressures. There are many advantages to using smaller particles, but they may not be essential for all separations. The most important advantages are higher resolution, faster analyses and increased sample load capacity. However, only the most demanding separations require the use of 3 μm packings. Many separation problems can be resolved with larger particle packings (e.g. 5 μm) that require less pressure. Flow-rate stability is another important pump feature that distinguishes pumps. Constant-flow systems are generally of two basic types: reciprocating piston and positive displacement (syringe) pumps. The basic advantage of both systems is that they produce highly repeatable elution volumes from run to run, regardless of viscosity, up to the pressure limit of the pump. Although syringe-type pumps have a pressure capability of up to 540 000 kPa (78 000 psi), they have a limited ability to form mobile phase gradients. Reciprocating piston pumps can maintain a liquid flow for an indefinite length of time, while a syringe pump needs to be refilled after the syringe volume has been displaced. Dual-headed reciprocating piston pumps provide more reproducible and pulse-free delivery of solvent, which reduces detector noise and enables more reliable integration of peak area. Reciprocating pumps now dominate the HPLC market and are even useful for micro-HPLC applications, as they can maintain a constant flow at flow rates in μL/min ranges.
An additional pump feature found on the more elaborate pumps is external electronic control. Although it adds to the expense of the pump, external electronic control is a very desirable feature when automation or electronic ally controlled mobile phase gradients are to be run. Alternatively, this becomes unnecessary when using isocratic methods. The degree of flow control also varies with pump expense. More expensive pumps include such state-of the-art technology as electronic feedback and multiheaded configurations.
Modern pumps have the following parameters:
• flow-rate range, 0.01–10 mL/min
• flow-rate stability, not more than 1% (short term)
• for size exclusion chromatography, flow-rate stability should be 0.2%
• maximum pressure, up to 34 500 kPa (5000 psi).