Evaluating column performance
Number of theoretical plates or column efficiency
The theoretical plate is an indirect measure of peak width at a specific retention time. Higher plate numbers indicate greater column efficiency and narrower peaks. The number of plates per metre of column (N) is calculated from either form of equation (18.4):

where tR is the time from injection to peak maximum for the solute, Wb is the peak width at base in units of time and Wh is the peak width at half height in units of time. Efficiency is a function of the column dimensions (diameter, length, film thickness or loading), the type of carrier gas and its flow, as well as the chemical nature of the solute and the stationary phase.